A) I, II, and III
B) II and III
C) I and II
D) I and III
Correct Answer
verified
Multiple Choice
A) I, II, and III
B) I and II
C) II and III
D) I and III
Correct Answer
verified
Multiple Choice
A) in the expiratory limb only, at a high point in the circuit
B) in both the expiratory and inspiratory limbs, at high points in the circuit
C) in the expiratory limb only, at a low point in the circuit
D) in both the expiratory and inspiratory limbs, at low points in the circuit
Correct Answer
verified
Multiple Choice
A) more water vapor it can hold
B) less water vapor it can hold
C) less efficient the humidifier is
D) more water vapor is lost
Correct Answer
verified
Multiple Choice
A) 30%
B) 50%
C) 70%
D) 90%
Correct Answer
verified
Multiple Choice
A) II, III, and IV
B) I, II, and III
C) III, IV, and V
D) I, II, III, IV, and V
Correct Answer
verified
Multiple Choice
A) membrane
B) bubble
C) heat-moisture exchanger
D) passover
Correct Answer
verified
Multiple Choice
A) 50% relative humidity (RH) at 20° to 22° C
B) 100% RH at 20° to 22° C
C) 80% RH at 28° to 32° C
D) 100% RH at 32° to 35° C
Correct Answer
verified
Multiple Choice
A) A variable air-entrainment port allows air mixing and different FIO2 values.
B) Liquid particles are generated by mechanical vibration energy.
C) They are pneumatically powered and attach directly to a flowmeter.
D) Baffling causes impaction or removal of large, unstable particles.
Correct Answer
verified
Multiple Choice
A) The gas at the airway is fully saturated.
B) The temperature setting of the heater is too high.
C) The airway temperature exceeds the dew point.
D) The temperature setting of the heater is too low.
Correct Answer
verified
Multiple Choice
A) failure of low-pressure alarms to detect disconnection
B) underhydration or impaction of secretions
C) hypoventilation due to increased dead space
D) unintended tracheal lavage from condensate
Correct Answer
verified
Multiple Choice
A) using an ultrasonic nebulizer instead of a jet nebulizer
B) using a 5% saline solution instead of a 3% concentration
C) having the patient rinse the mouth or blow the nose before induction
D) using the lowest possible aerosol density (high flow and low output)
Correct Answer
verified
Multiple Choice
A) Of the exhaled water vapor, 80% returns to the patient on inspiration.
B) The inspired temperature is 80% of the expired temperature.
C) Of the exhaled water vapor, 20% returns to the patient on inspiration.
D) The device provides 80% relative humidity at body temperature.
Correct Answer
verified
Multiple Choice
A) at least every shift
B) when contaminated by secretions
C) at least every day
D) when condensate is visible
Correct Answer
verified
Multiple Choice
A) 40 mg/L water vapor exhaled; 27 mg/L returned
B) 40 mg/L water vapor exhaled; 20 mg/L returned
C) 27 mg/L water vapor exhaled; 40 mg/L returned
D) 44 mg/L water vapor exhaled; 37 mg/L returned
Correct Answer
verified
Multiple Choice
A) aerosolization of condensate during disconnection
B) underhydration or impaction of secretions
C) hypoventilation due to increased dead space
D) unintended tracheal lavage from condensate
Correct Answer
verified
Multiple Choice
A) warm and humidify gases delivered to the trachea via ventilator circuits
B) humidify therapeutic gases delivered at high flows to the lower airway
C) provide extra humidity for a patient with thick or retained secretions
D) warm and humidify dry therapeutic gases delivered to the upper airway
Correct Answer
verified
Multiple Choice
A) I, II, and III
B) I and II
C) II and III
D) I and III
Correct Answer
verified
Multiple Choice
A) I and II
B) II and III
C) I and III
D) I, II, and III
Correct Answer
verified
Multiple Choice
A) I, II, and III
B) II and III
C) II
D) III
Correct Answer
verified
Showing 21 - 40 of 112
Related Exams